Ambetronics

Biogas Analyzer Panel – BIO-400-S-PANEL

The BIO-400-S-PANEL Biogas Analyzer is a cutting-edge solution for the precise measurement and analysis of biogas in various applications. Developed with infrared sensor technology in India, this analyzer boasts a digital display featuring CH4, CO2, O2, and H2S concentrations on a graphic LCD display with adjustable backlight. Designed for versatility, the optional Dew Point sensor allows monitoring of NH3, CO, and H2 gases as needed. The plug-in sensor offers a full year of guaranteed protection against hazardous gases, with calibration due alerts and indications for user-friendly operation. The menu is password-protected, and calibration is easily performed using front keys. Equipped with low and high configurable alarms with LED indicators, the analyzer ensures safety. The software allows calibration through front keys, and data logging is facilitated with a substantial capacity of 100,000 records, visible in ASCII format when downloaded to a PC via USB interface (in Excel format).

1. Innovative Infrared Sensor Technology: Developed in India, the biogas analyzer employs advanced infrared sensor technology for precise and reliable gas measurement.
2. Digital LCD Display: The biogas analyzer features a 4×2 LCD display with adjustable backlight, providing a clear presentation of CH4, CO2, O2, and H2S concentrations for quick and easy monitoring.
3. Optional Dew Point Sensor: Offers flexibility with an optional dew point sensor for monitoring CO gas as per specific application requirements.
4. Plug-in Sensor with Longevity: The plug-in sensor ensures a full year of warranted protection against hazardous gases, enhancing the longevity and reliability of the analyzer.
5. Calibration Due Indication: User-friendly calibration due indication and alert system, ensuring the accuracy and reliability of gas measurements over time.
6. User-Friendly Interface: Easy programming with just three keys on the front panel (SET, SHIFT/ACK, INCR), making it accessible for users with varying levels of expertise.
7. Password-Protected Function Menu: Enhanced security with a password-protected function menu, preventing unauthorized access and adjustments.
8. Configurable Alarms: Offers low and high configurable alarms with LED indications, providing visual alerts for quick response to potentially hazardous gas levels.
9. Software Calibration: Calibration is simplified with software calibration performed entirely using front keys, streamlining the maintenance process.
10. Data Logging Facility: The biogas analyzer supports data logging with an impressive capacity of 100,000 records. Data is easily accessible in ASCII format when downloaded to a PC via USB interface (in Excel format).
11. Self-Purging Technology: Incorporates self-purging technology activated by a simple two-key press, ensuring the accuracy of subsequent gas measurements.
12. Optional Calibration Kit: The availability of an optional calibration kit provides users with the tools needed for periodic maintenance and calibration.

Technical Specifications:

1. Detection Method: Suction/pressurized, depending on the application, ensuring adaptability to different environments.
2. Display: 3-key LCD with settable backlight for optimal visibility in various lighting conditions.
3. Set Point: Two independent set points (AL1 & AL2) with LED indication for efficient monitoring and control.
4. Response Time: Rapid response time of less than 30 seconds for timely detection of gas concentrations.
5. Warm-up Time: Quick warm-up time of 60 seconds, minimizing downtime during operation.
6. Electrical Specifications:
– Supply Voltage: 230 VAC ± 10%, providing stable and reliable power for continuous operation.
7. Output Signals:
– Download Capability: 100,000 records logging capacity with data visible in ASCII format during PC download through USB interface or RS-485 MODBUS RTU.
– Setting Indication: Over/Open range indication on the LCD for ease of setup and monitoring.
– Alarm Outputs: Two relay outputs for low and high alarms, along with RS-485 MODBUS RTU output for integration with control systems.
8. Environmental Specifications:
– Operating Temperature: Wide operating temperature range of 0 to 50°C, ensuring suitability for diverse environmental conditions.
– Storage Temperature: 0 to 60°C, allowing for safe storage when not in use.
– Humidity: Below 95% RH, non-condensing, with a recommended range of 40% to 60% RH for optimal performance.
– Atmospheric Pressure: Operates within the atmospheric pressure range of 700 mbar to 1200 mbar.
9. Free Accessories:
– Operating Manual: Comprehensive manual for easy setup, operation, and maintenance.
– Built-in Sample Pump and Accessories: Includes a 2-way valve, filters, flow meter, and solenoid valve, providing essential components for efficient gas sampling and analysis.
10. Enclosure Details:
– Cabinet Material: Sturdy MS (Mild Steel) construction for durability and protection of internal components.
– Size (External): Compact dimensions of 600 mm (L) x 600 mm (H) x 400 mm (D), ensuring versatility in installation.
– Weight: Maximum weight of 40 kg, facilitating easy handling and installation.

Biogas Plants
Farm Digester Gas Monitoring
Food Processing Biogas Monitoring
Wastewater Biogas Monitoring
Methane Recovery

Get Quote From Our Sales Team